Mini Samples ELISA Kit for Melatonin (MT)

Cloud-Clone Corp.
Product Code: MEA908Ge
Product Group: ELISA Kits
CodeSizePrice
MEA908Ge-24T24T£285.00
Quantity:
MEA908Ge-48T48T£475.00
Quantity:
MEA908Ge-96T96T£666.00
Quantity:
MEA908Ge-96T*596T*5£2,889.00
Quantity:
MEA908Ge-96T*1096T*10£5,429.00
Quantity:
Prices exclude any Taxes / VAT

Overview

Regulatory Status: RUO
Application: Enzyme-Linked Immunosorbent Assay (ELISA)

Images

1 / 1

Further Information

Alternative Names:
N-Acetyl-5-Methoxytryptamine
Assay length:
3h
Assay Procedure Summary:
1. Prepare all reagents, samples and standards;
2. Add 25µL standard or sample to each well.
    And then add 25uL prepared Detection Reagent A immediately.
    Shake and mix. Incubate 1 hour at 37°C;
3. Aspirate and wash 3 times;
4. Add 50µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
5. Aspirate and wash 5 times;
6. Add 50µL Substrate Solution. Incubate 10-20 minutes at 37°C;
7. Add 25µL Stop Solution. Read at 450 nm immediately.
Detection range:
12.35-1,000pg/mL
Format:
48T, 96T, 96T?5, 96T?10, 96T?100
Item Name:
Melatonin
Method:
Competitive Inhibition
Precision:
Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Mini Samples Melatonin (MT) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Mini Samples Melatonin (MT) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%
Research Area:
Metabolic pathway;Tumor immunity;Neuro science;Hormone metabolism;
Sample type:
Serum, plasma and other biological fluids
Sensitivity:
The minimum detectable dose of this kit is typically less than 5.31pg/mL
Specificity:
This assay has high sensitivity and excellent specificity for detection of Mini Samples Melatonin (MT).
No significant cross-reactivity or interference between Mini Samples Melatonin (MT) and analogues was observed.
Stability:
The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.
Test Principle:
This assay employs the competitive inhibition enzyme immunoassay technique. A monoclonal antibody specific to Mini Samples Melatonin (MT) has been pre-coated onto a microplate. A competitive inhibition reaction is launched between biotin labeled Mini Samples Melatonin (MT) and unlabeled Mini Samples Melatonin (MT) (Standards or samples) with the pre-coated antibody specific to Mini Samples Melatonin (MT). After incubation the unbound conjugate is washed off. Next, avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. The amount of bound HRP conjugate is reverse proportional to the concentration of Mini Samples Melatonin (MT) in the sample. After addition of the substrate solution, the intensity of color developed is reverse proportional to the concentration of Mini Samples Melatonin (MT) in the sample.