Overview

Regulatory Status: RUO

Documents

Further Information

Alternative Names:
CK-MM; CKMM; CK3; CK-M; MM-CK; M-CK; Creatine kinase M-type, N-terminally processed
Assay Length:
2h, 40min
Assay Procedure Summary:
1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 2 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 100µL Substrate Solution. Incubate 10 minutes at 37°C;
8. Read RLU value immediately.
Detection Range:
0.14-100ng/mL
Item Name:
Creatine Kinase, Muscle
Method:
Double-antibody Sandwich
Precision:
Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Creatine Kinase, Muscle (CKM) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Creatine Kinase, Muscle (CKM) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%
Sample Type:
serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Sensitivity:
The minimum detectable dose of this kit is typically less than 0.05ng/mL
Specificity:
This assay has high sensitivity and excellent specificity for detection of Creatine Kinase, Muscle (CKM).
No significant cross-reactivity or interference between Creatine Kinase, Muscle (CKM) and analogues was observed.
Stability:
The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.
Test Principle:
The microplate provided in this kit has been pre-coated with an antibody specific to Creatine Kinase, Muscle (CKM). Standards or samples are then added to the appropriate microplate wells with a biotin-conjugated antibody specific to Creatine Kinase, Muscle (CKM). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. Then the mixture of substrate A and B is added to generate glow light emission kinetics. Upon plate development, the intensity of the emitted light is proportional to the Creatine Kinase, Muscle (CKM) level in the sample or standard.;
Uniprot ID:
P00564