ELISA Kit for TEK Tyrosine Kinase, Endothelial (Tie2)

Cloud-Clone Corp.
Product Code: SEA126Ra
Product Group: ELISA Kits
CodeSizePrice
SEA126Ra-24T24T£213.00
Quantity:
SEA126Ra-48T48T£349.00
Quantity:
SEA126Ra-96T96T£485.00
Quantity:
SEA126Ra-96T*596T*5£2,077.00
Quantity:
SEA126Ra-96T*1096T*10£3,897.00
Quantity:
Prices exclude any Taxes / VAT

Overview

Regulatory Status: RUO
Target Species: Rat
Application: Enzyme-Linked Immunosorbent Assay (ELISA)

Images

1 / 1

Further Information

Alternative Names:
CD202b; CD202-B; TEK; VMCM1; Tyrosine Kinase With Immunoglobulin Like And EGF Like Domains 2; Venous Malformations Multiple Cutaneous And Mucosal
Assay length:
3h
Assay Procedure Summary:
1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 2 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.
Detection range:
0.312-20ng/mL
Format:
48T, 96T, 96T?5, 96T?10, 96T?100
Item Name:
TEK Tyrosine Kinase, Endothelial
Method:
Double-antibody Sandwich
Precision:
Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level TEK Tyrosine Kinase, Endothelial (Tie2) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level TEK Tyrosine Kinase, Endothelial (Tie2) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%
Research Area:
Enzyme & Kinase;Tumor immunity;Infection immunity;
Sample type:
Serum, plasma, tissue homogenates and other biological fluids
Sensitivity:
The minimum detectable dose of this kit is typically less than 0.135ng/mL
Specificity:
This assay has high sensitivity and excellent specificity for detection of TEK Tyrosine Kinase, Endothelial (Tie2).
No significant cross-reactivity or interference between TEK Tyrosine Kinase, Endothelial (Tie2) and analogues was observed.
Stability:
The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.
Test Principle:
The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to TEK Tyrosine Kinase, Endothelial (Tie2). Standards or samples are then added to the appropriate microtiter plate wells with a biotin-conjugated antibody specific to TEK Tyrosine Kinase, Endothelial (Tie2). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain TEK Tyrosine Kinase, Endothelial (Tie2), biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of TEK Tyrosine Kinase, Endothelial (Tie2) in the samples is then determined by comparing the O.D. of the samples to the standard curve.
Uniport ID:
D3ZCD0

References