High Sensitive ELISA Kit for Prothrombin Fragment 1+2 (F1+2)

Cloud-Clone Corp.
Product Code: HEA710Hu
Product Group: ELISA Kits
CodeSizePrice
HEA710Hu-24T24T£217.00
Quantity:
HEA710Hu-48T48T£356.00
Quantity:
HEA710Hu-96T96T£496.00
Quantity:
HEA710Hu-96T*596T*5£2,125.00
Quantity:
HEA710Hu-96T*1096T*10£3,987.00
Quantity:
Prices exclude any Taxes / VAT

Overview

Regulatory Status: RUO
Target Species: Human
Application: Enzyme-Linked Immunosorbent Assay (ELISA)

Images

1 / 1

Further Information

Assay length:
3h
Assay Procedure Summary:
1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 2 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.
Detection range:
0.11-80pg/mL
Format:
48T, 96T, 96T?5, 96T?10, 96T?100
Item Name:
Prothrombin Fragment 1+2
Method:
Double-antibody Sandwich
Precision:
Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level High Sensitive Prothrombin Fragment 1+2 (F1+2) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level High Sensitive Prothrombin Fragment 1+2 (F1+2) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%
Sample type:
serum, plasma and other biological fluids
Sensitivity:
The minimum detectable dose of this kit is typically less than 0.04pg/mL.
Specificity:
This assay has high sensitivity and excellent specificity for detection of High Sensitive Prothrombin Fragment 1+2 (F1+2).
No significant cross-reactivity or interference between High Sensitive Prothrombin Fragment 1+2 (F1+2) and analogues was observed.
Stability:
The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.
Test Principle:
The microplate provided in this kit has been pre-coated with an antibody specific to F1. Standards or samples are then added to the appropriate microplate wells with a biotin-conjugated antibody specific to F2. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain F1+2, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ? 10nm. The concentration of F1+2 in the samples is then determined by comparing the O.D. of the samples to the standard curve.
Uniport ID:
P00734